Medima Syringe Infusion Pumps

Syringe pump

 

Medima S

Basic model, extremely simple to use. Operates in: ml/min, ml/h and ml/24h.

Calculates flow rate basing on volume and delivery time.

In addition, can store in memory up to 120 drug names. Displays one throughout infusion.

 

Medima S1

Infusion flow rate programmable in ml/h and other frequently used units including: ng, µg, mg, µU, mU, U, kU, µmol, mmol, mol, / kg / min, h, 24 h.

Numerous infusion modes such as: continuous infusion, intermittent infusion, infusion with profile (up to 24 cycles).

In addition, can store in memory a drug library of 120 full drug dosing protocols, but without infusion parameter limits. Displays drug name throughout infusion.

Comprehensive configuration options: advanced protection system, night mode, alarm options.

 

Medima S2

All options of Medima S1 plus drug library with limits applicable to all infusion parameters.

Soft limits (min max) warn when recommended values are exceeded but do not prohibit the introduction of changes. Hard limits (min max) disable introduction of any values from outside prescribed range.

 

Medima S-PCA

Designed to fight against acute pain. Equipped with a patient-operated button which starts the delivery of programmed drug dose and a syringe cover with a lock.

PCA

An option to switch to standard infusion modes same as for model S2.

All options of Medima S2 including drug library with infusion parameter limits.

Versions with or without syringe cover.

^ back to top

 

Basic functions

All Medina pumps are operated in the same manner (differences between models considered) and have a number of common features:

Large and legible graphic display which constantly displays:

  • Name of the medicine
  • Flow rate
  • Power status and power supply type
  • Infusion pressure and preset occlusion alarm level
  • Size and type of syringe

Syringe Pump Display

Wide range of syringes can be selected from all major brands of the following sizes: 5, 10, 20, 30, 50 ml.

Automatic syringe loading and recognition simplifying and speeding up pump maintenance, reducing the initial infusion errors.

Simple to set up and easy to operate. Quick data entry with the use of numerical keyboard. All parameters can be adjusted without infusion interruption.

Different infusion units ng, µg, mg, µU, mU, U, kU, µmol, mmol, mol, / kg / min / hour / 24h.

Data lock. Programmed parameters can be password-protected. Additionally sophisticated function protection system allows blocking access to chosen functions, i.e. infusion start, bolus start, occlusion threshold change, pump switching-off.

Bolus fully automatic and manual. Possibility of programming dose, time or flow rate.

Automatic reduction of residual bolus after occlusion release.

6 levels of occlusion pressure. Possibility to adjust the pressure level during infusion. Pressure indication displayed constantly during delivery.

Sophisticated system of alarms. Allows different tones and loudness. Built in night mode with the option to set lower sound level as well as reduce display brightness.

Event log. Allows storing full infusion history (parameters, operator activities and alarms along with the date and time of occurrence). The stored information can be viewed either on the pump's display or on PC.

Drug library. Up to 120 drug dosing procedures can be stored in pump's memory along with all the essential infusion parameters. Additionally upper and lower, soft and hard limits for all infusion parameters can be entered in a drug library to minimize medications errors and ensure overdose protection. Soft limits warn about exceeding recommended values, but do not block values entered. Hard limits prevent from entering values that are beyond the range. The drug library can be entered as follows:

  • directly from the pump keyboard
  • sent from pump to pump by RS232 interface
  • created in Medima Drug Editor software and set to the pump directly from a PC or via MedimaNet

^ back to top

 

Technical data

Flow-rate range

0.1 - 2000 ml/h for 50 ml syringe

0.1 - 400 ml/h for 5 ml syringe, programmable in 0.1 ml/h increments

Volume of infusion

0.1 - 9999 ml, in 0.1 ml increments

Infusion time

1 min - 200 h

Bolus rate

Up to 2000 ml/h for 50 ml syringe

Up to 400 ml/h for 5 ml syringe, programmable in 1 ml/h increments

Programmable volume (dose), time or rate

Automatic and manual bolus

Induction dose

Up to 2000 ml/h for 50 ml syringe

Up to 400 ml/h for 5 ml syringe, programmable in 1 ml/h increments

Programmable volume (dose), time or rate

KVO

Programmable, 0 - 5.0 ml/h

Flow rate accuracy

Technical accuracy ±1%, ±2% according to EN 60601-2-24 *

Syringe types

5, 10, 20, 30/35, 50/60 ml by numerous manufacturers **,
automatic recognition of syringe

Occlusion pressure

6 levels, 0.2 - 1.2 bar (kg/cm2), 150 - 900 mmHg,
automatic reduction of residual bolus after occlusion release

Graphic presentation of dosing pressure and the set occlusion level (pictograms)

Operation modes

Model S

Model S1

Model S2

Model S-PCA

ml/h, ml/min, ml/24h, volume over time

-

ng, µg, mg, µU, mU, U, kU, nmol, µmol, mmol, mol
/ kg / min, hour, 24h

-

intermittent infusion

-

profile - 24 cycles,
ramp on / ramp off (TPN),

-

-

-

PCA mode

Additional functions

Automatic syringe fixation

Possibility to enter hard and soft limits for all infusion parameters - Models S2 and S-PCA

Drug library up to 120 programmable drugs (Models S1, S2 and S-PCA)

Drug library and configuration pump to pump transfer

Programmable infusion pause; 1 min up to 24h

Delayed start; 1 min up to 24h

Programmed parameters and chosen functions protected by password

Alarm volume adjustment and choice of sound type

Night mode; reduced display brightness; programmable sound loudness, programmable time of start and end of infusion

Simple connection / disconnection to docking station (during infusion)

Advanced system of pump testing

Test of battery wear out

Possibility to connect into hospital information system

Open system: software update via RS232C

Alarms

No mains, low battery, battery discharged, 1÷30 min to the end of infusion, end of infusion, 1÷30 min to the end of syringe, empty syringe, pressure increase, occlusion, line disconnected (rapid pressure drop), incorrect syringe, pause, end of pause in infusion, open cover (in S-PCA), internal malfunction

Event log

2000 records of operation activities or alarms with date and time of event

Power supply

100-240 VAC, 50/60 Hz, max 15W, 12 - 15 VDC, max. 1 A

Battery

NiMH, battery life: ≥ 20 h / 5 ml/h

Recharging time: max. 5 hours

Easy battery access

Interface

RS232C and IR link with the docking station

Alarm connection

24V, 1 A, NO / NC

Dimensions

355 x 115 x 166 mm (excluding pole clamp)

Weight

2,4 kg

Operating conditions

Temperature 5 - 40°C, humidity max. 90%

Classification

CF, protection class I, defibrilation proof, IP22

Compiles with safety standards

EN60601-1, EN60601-1-2, EN60601-2-24, EN1789,
MDD93/42/EEC - II B

Accessories

Connecting lead RS232C

Connecting lead 12 V

Connecting lead for staff call

PCA syringe cover

Software package Medima ToolBox

* Technical accuracy is 1%. Flow rate accuracy depends on the quality of syringe.
** List of syringes available on request. There is a possibility to calibrate the pump for other syringes according to customer wish.

Out to concern for quality improvement and due to changing standards the information given may be subject to change. Contact Medima Ltd to update the information.

^ back to top

CE1011

  x    
      x      
MEDIMA - infusion pumps - syringe, volumetric (peristaltic) and PCA pumps